Custom Part Manufacturing Guide
Calculating Kinematics of a CTR
Graph Paper In-Plane Bending Experiment
Out-of-Plane Rotation Experiment
The actuation system is designed such that tubes can be inserted and removed in under 3 minutes. They use a twist and lock mechanism to allow for both easy access and rigid control.
Assemble your concentric tubes such that they are nested properly within each other.
Separate the carts in the actuation system by about 50mm. This will allow you to reach in between carts to lock the tubes in place.

Insert the tubes from the front of the actuation system. You will need to twist the tubes to allow the cubes to pass through the carts.

Insert the innermost tube cube into the last cart. Do not push it all of the way through. Rotate the cube by about $45^\circ$. Insert on the tube locks to hold the tube in place. You should be able to gently tug on the tube without it moving. If the tube moves at all, this means it did not properly lock. You most likely inserted the cube to far into the cart.